SciAps Z-901 LIBS Analyzer
The Z-901 for alloy analysis measures 15-20 elements in 7 common alloy bases, including aluminum, stainless, iron, copper, titanium, nickel and cobalt bases. Other bases may be added for a small fee, or users may add their own calibrations.
Alloy: Argon-purge for high precision and improved detection limits. Many customers use argon-purge to analyze and sell mill-ready aluminum scrap. Quantify low ppm levels of beryllium and boron.
QuickSort: Air-burn for rapid material sorting or grade identification.
Features
– LIBS Sensor – advanced spectrometer design for high resolution and wide range
– Internal Camera – precise targeting of analysis location
– Macro Camera – photo documentation of samples, reading barcodes and QR codes
– Report Generation – full-featured, with available cloud data management and reporting
– Narrow Snout – tapered for welds or difficult-to-access test locations
– Laser Safety Sensor – patented sample sensor allow Class 1 operation, subject to LSO approval
– Intuitive Android – operating system, with app based software
– High Resolution Display – rear-facing display for easy viewing
Technical Data
– Spectral Range: 200-420nm
– Processing Electronics: ARM Quad Cortex -A53 1.2GHz memory: 2 GB LPDDR3, 16 GB eMMC
– Data Storage: Results Storage: 32 GB SD
– Dimensions: 10.75 x 2.875 x 8.625 inches
– Display: 2.7-inch high-brightness color touchscreen, readable in all conditions. Rear-facing display for easy results viewing
– Power: On board rechargeable Li-ion battery, rechargeable inside device or with external charger, AC power
– Connectivity: Built on Google’s Android platform for real-time data exporting, including built-in WiFi (IEEE 802.11b/g/n), Bluetooth (BR/EDR+BLE), GPS and USB-C to connect to virtually any information management system
– Security: Password protected. Multi-user support with configurable access
– Regulatory: CE, RoHS, USFDA registered. Class 3b laser. Sample sensor on board, allows for operation under Class 1 conditions subject to local LSO approval
– Sample viewing: Integrated camera and laser target indicator before and during analysis provide proper sample alignment. Includes 2nd “macro camera” for scanning QR or barcodes and for photo-documentation and report generation
Laser Raster: On-board Y stage for rastering laser to discrete locations for targeted analysis or averaging
– Atmosphere: SciAps proprietary Opti-Purge inert argon environment improves spectral signal-to-noise ratio; improves performance in UV range
– Calibration Check: Internal 316 stainless check standard for calibration verification and wavelength scale validation
– Drift Connection: On-board automated drift correction software with factory or user-provided reference materials
Grade Library: 500+ grades, multi-library support, libraries may be added or edited